Home Products Contact Us Disclaimer

Specifications


Previous level

 

Specifications of Abacus Junior

&

Abacus Junior Vet

 

Sample volume:                       15-25 µl whole blood

Aperture diameter:                    80 µm (in one universal measuring chamber)

Measured parameters:             18 parameters, including 3-part WBC differential: WBC, LYM, MID, GRA, LYM%, MID%, GRA%, RBC, HCT, MCV, RDW, HGB, MCH, MCHC, PLT, PCT, MPV, PDW

Throughput:                              Human model: 25 tests/hour

                                               Vet model: 20 tests/hour

Characteristics:

Accuracy
max deviance from expected

Reproducibility

(CV)

Carry over
between samples

Test range

 

WBC

3%

< 3%

< 1%

4.0 – 20.0

 

RBC

3%

< 3%

< 1 %

4.00 – 15.00

 

HCT

3%

< 3%

< 1 %

25.0 – 50.0

 

MCV

2%

< 1%

N/A

50 – 90

 

HGB

2%

< 2%

< 1 %

9 – 16

 

PLT

5%

< 5%

< 3 %

200 – 900

           

Sampling method:                    Open tube system with automatic sample rotor

Sample types:                        Human model: Human (general), Male, Female, Baby, Toddler and Child

                                               Vet model: Dog, Cat, Horse, plus 6 user-defined

Fault statistics:                         RBC/WBC clogging < 1% of analyses (normal use)

Clog prevention:                       High-voltage pulse and chemical cleaning of the aperture in each analysis cycle using Cleaner  reagent

Cleaning procedure:                High-voltage burst of aperture, high-pressure back-flush, chemical cleaning of aperture using Cleaner

Quality control:                         6  levels, including: mean, ± range, SD and CV for all measured and calculated parameters, 16- and 64-day Levey-Jennings charts, separate QC database

Calibration :                            Automatic based on 1 or 3 measurements, or manual calibration of WBC, HGB, RBC, PLT, MCV, RDW, MPV (separate factors for pre-diluted and Vet modes), monitoring of calibration  factors by calibration events

Multi-user feature:                    3-level multi-user operation with selective privilege levels, user identification with ID and password

User interface:                        Easy-to-use, menu driven user interface with 6 software buttons (with icons), 6 hardware function buttons (above LCD), cursor and numeric keys

Languages available:               Chinese, English, French, German, Greek, Indonesian, Italian, Persian, Polish, Portuguese, Romanian, Russian, Spanish, Vietnamese

Data storing capacity:                1000 results, including histograms

Host computer interface:            Serial (RS-232) computer link

Data back-up Interface:             3.5” floppy disk

Software upgrade:                     3.5” floppy disk or slim CD-ROM (optional)

Printer interface:                       Centronics (parallel) or USB (optional)

Optional built-in printer:             “Easy Paper Operation” built-in thermal printer

Display:                                   240 x 128 dots high-contrast, backlit, graphic LCD

Keypad:                                   29 foil keys + START button

External keyboard:                   Standard PS/2 compatible keyboard

Power supply:                          Universal, external 12VDC, 8A power module

Power supply (input):                90-260V, 50-60Hz, 10W stand by, 80W max.

Operating temperature:            15-35°C

Dimension (WxDxH):                320 x 260 x 365 mm

Net weight:                             12 kg

 

 

 

 

Copyright © 2004 ANJOTECH. All rights reserved.