Specifications of Abacus
Junior
&
Abacus Junior Vet
Sample volume:
15-25 µl whole blood
Aperture diameter:
80 µm (in one universal measuring chamber)
Measured parameters:
18 parameters, including 3-part WBC differential: WBC, LYM, MID, GRA, LYM%,
MID%, GRA%, RBC, HCT, MCV, RDW, HGB, MCH, MCHC, PLT, PCT, MPV, PDW
Throughput:
Human model:
25 tests/hour
Vet model:
20 tests/hour
|
Characteristics: |
|
|
Carry over
between samples |
Test range |
|
|
WBC |
|
|
|
|
|
|
RBC |
3% |
< 3% |
< 1 % |
4.00 – 15.00 |
|
|
HCT |
3% |
< 3% |
< 1 % |
25.0 – 50.0 |
|
|
MCV |
2% |
< 1% |
N/A |
50 – 90 |
|
|
HGB |
2% |
< 2% |
< 1 % |
9 – 16 |
|
|
PLT |
5% |
< 5% |
< 3 % |
200 – 900 |
| |
|
|
|
|
|
Sampling method:
Open tube system
with automatic sample rotor
Sample types:
Human model: Human (general), Male, Female, Baby, Toddler and
Child
Vet model:
Dog, Cat, Horse, plus 6 user-defined
Fault statistics:
RBC/WBC clogging < 1%
of analyses (normal use)
Clog prevention:
High-voltage pulse
and chemical cleaning of the aperture in each analysis cycle using Cleaner
reagent
Cleaning procedure:
High-voltage burst
of aperture, high-pressure back-flush, chemical cleaning of aperture using
Cleaner
Quality control:
6
levels, including: mean, ± range, SD and CV
for all measured and calculated parameters, 16- and 64-day Levey-Jennings
charts, separate QC database
Calibration
:
Automatic based on 1 or 3 measurements, or
manual calibration of WBC, HGB, RBC, PLT, MCV, RDW, MPV (separate factors for
pre-diluted and Vet modes), monitoring of calibration
factors by calibration events
Multi-user feature:
3-level multi-user
operation with selective privilege levels, user identification with ID and
password
User interface:
Easy-to-use, menu driven user
interface with 6 software buttons (with icons), 6 hardware function buttons
(above LCD), cursor and numeric keys
Languages available:
Chinese, English, French, German, Greek, Indonesian, Italian, Persian, Polish,
Portuguese, Romanian, Russian, Spanish, Vietnamese
Data storing capacity:
1000
results, including histograms
Host computer interface:
Serial (RS-232) computer link
Data back-up Interface:
3.5” floppy disk
Software upgrade:
3.5” floppy disk or slim CD-ROM (optional)
Printer interface:
Centronics (parallel) or USB (optional)
Optional built-in printer:
“Easy Paper Operation” built-in thermal printer
Display:
240 x 128 dots
high-contrast, backlit, graphic LCD
Keypad:
29 foil keys + START button
External keyboard:
Standard PS/2 compatible keyboard
Power supply:
Universal, external 12VDC, 8A power module
Power supply (input):
90-260V,
50-60Hz, 10W stand by, 80W max.
Operating temperature:
15-35°C
Dimension (WxDxH):
320 x 260 x
365 mm
Net weight:
12 kg
|